Tag: siemens

  • NIST 1976c custom mount

    NIST 1976c custom mount

    It’s easy to forget how much of our scientific work hinges upon comparative data. The entire field of metrology is concerned with the verification and maintenance of “standard reference materials” (SRMs). Creating a perfect reference standard essentially involves proving a negative. In the XRD world, we need to prove that there are no impurities, no crystalline defects, no unaccounted for thermal variations, no stress/strain effects present, and above all, that the first unit produced is effectively identical to the last and all between.

    There really isn’t any one material that checks all the boxes, but the NIST gets very close thanks to the efforts of Mr. Jim Cline and his associates. The hardware they’re using is completely custom to the point that it bares no resemblance to the instrumentation we normally work with. There’s a great page on their divergent-beam lab here.

    Image of DBD

    The NIST 1976 material has been a mainstay in regular monitoring and certification of XRD system performance for many years and is now on its third (c) generation. With this most recent revision the shape of the standard has been changed from a flat plate to a round disk. This allows for much greater compatibility across the range of sample holders in the market.

    This week we made a custom mount for a company using a Siemens D5005 with a 40-position autosampler. It’s common to mount these, but this is the first time I can recall doing so for this particular autosampler so I thought it was worth sharing a little about the material and a picture of the finished product.

  • What’s really coming out of my XRD tube?

    What’s really coming out of my XRD tube?

    We’ve been experimenting with better ways to quantify the quality of XRD tubes in the shop. We use these tests on new and used tubes to monitor performance in two key areas. 1) Intensity 2) Spectral purity.

    What we’ve settled on is a test that involves a wavelength-dispersive approach which gives us a lot of intensity to work with while eliminating background scatter and fluorescence effects. Basically, we’re able to extract more information from the data because the “noise” is almost zero.

    We used Jade Pro to evaluate the scans, but they’re not D-spacing vs intensity as one would normally expect. This scan represents Wavelength vs intensity more like one would see in a WDXRF spectrometer. Cu KA1 and 2 are obvious, as is Cu KB1. Many of the current generation of XRD users have never seen a W LA1 peak in their data, but it’s clearly visible here as this is an older tube. What I’ve never been able to see before is the W La2 peak in the green scan. You’re looking at a peak that is ~62eV separated from W La1. No XRD detector on the market has energy resolution like that so these would always be lumped together so you’d see a series of additional peaks from every d-spacing in the sample in the diffractogram. Only a handful of detectors (our SDD-150 for example) could even separate the W La from the Cu K lines. That’s the power of wavelength-dispersive techniques. Incidentally, the most common device for cleaning up superfluous energy emissions in XRD data is a diffracted-beam monochromator and they eliminate all the W La through a secondary diffraction event much like what we’re doing.

    Characterizing emissions is nothing new. In fact, I started wanting to improve this after listening to a talk at DXC about Jim Cline’s famous XRD system which is used at NIST to perform the primary data collection on the CRMs we all use. To paraphrase Sir Arthur Conan Doyle, “When you explain every extraneous data point, the remaining information is the pure truth of the sample”.

  • Fully digital autosampler install

    Fully digital autosampler install

    Another fully-rebuilt, digital autosampler out in the wild. This one is on a system that already has one of our Si-Drift Detectors and an awesome ICDD Jade Pro/PDF-4+ software package. We’ve got all the fancy new hardware at our in-house lab, but when we need the absolute best data, this is our goto configuration.